| Internal sampling rates: |
1GH max to 100Hz min |
| No. of channels: |
16 |
| Data buffer: |
1Meg samples/channel for 500MHz and below sampling
20K samples/channel for 1GHz sampling |
| Input bandwidth: |
100MHz |
| Input impedance: |
1MΩ//5pF |
| Threshold voltage: |
-5V to +5V |
| Input voltage (absolute limits): |
-60V to +60V |
| External clock |
Synchronized capture: 50 MHz Max
Synchronized capture into linear/ring buffer.
Ring buffer configuration: Operates with pre and post trigger buffer
Linear capture may start on trigger condition or immediately |
| Digital logger: |
1 Second to 1-hour sampling rates. |
| Pre-trigger and post-trigger buffer setting: |
The data buffer is divided in pre- and post trigger sections. The pre-/post-trigger buffer relation may be changed in 1000 samples steps. |
| Extended Capture Time |
Dependant on the sampling rate. At 500Mhz: 33ms Max, at 1.25KHz:1677s max |
| Internal sampling clock stability |
100 PPM |
| Instrument integration |
Can combine with pattern generator functionality |
|
| (Pattern valid for any duration) |
1,0, and "don't care" ("X") conditions selectable on all channels. |
| Pattern valid < Duration |
Glitch Capture. Pattern duration may be specified in steps of 20ns to 1 ms. Duration tolerance of ±20ns |
| Pattern valid > Duration |
Pattern duration may be specified in steps of 20ns to 1ms. Duration tolerance of ±20ns |
|
| Triggers On: |
Rising edge, falling edge, either rising or falling edge (change of state), of any one or combination of channels |
| Mouse/Keyboard |
Trigger may be forced. |
Trigger Methods: Combination Pattern/Edge
|
| Edge OR pattern |
Edge or pattern condition that occurs first will cause a trigger. |
| Edge AND Pattern |
For a trigger the edge condition must occur while the pattern condition is valid. |
| Edge THEN Pattern |
The first condition is required before the second condition. At least 30ns is required from detection of the first condition to the second. |
| Application of trigger conditions |
| Single capture |
Captures a single set of data |
| Continuous capture display: |
Unconditionally Continuous: Trigger is forced internally and display updated with regular intervals.
Conditionally Continuous: Display is updated when a trigger condition is detected. Any of the above trigger conditions may be set as described above.The display is updated on detection of a specified trigger condition. |
| Supported Operating Systems: |
Microsoft Windows 98,ME,2000,XP,Vista32 or later compatible versions. |
| PC |
Data capture speed is unaffected by PC- speed
Ease of use: The software is very easy to use. Most functions are directly selectable by means of function buttons on the main scree |
| Multitasking environment |
Run the logic analyzer software and user’s unit under test (UUT) software simultaneously (if the UUT is PC based).
Hardware that interfaces with the PC may be driven from the PC while real-time measurements are performed on the hardware. (The logic analyzer and UUT software are simultaneously run on the same PC). The analyzer and the user program may be viewed simultaneously in the Windows environment. |
| Ease of use: |
The software is very easy to use. Most functions are directly selectable by means of function buttons on the main screen. |
| User UUT application program launching: |
The user application program may be run from the logic analyzer program menu or from Windows. |
| |
|
|
| No. of channels: |
Any number of channels may be displayed. |
| Channel/Group names |
User specified signal/group names |
| Display order: |
Channel display order user specified |
| Display Colors |
Specify colors for individual signals, background. |
| Zooming: |
Data zoom: Zoom in/out/previous/all/between cursors/etc
All captured data may be viewed on one screen.
Zooming in to only a few samples displayed on the screen
Easy data display window sizing using the mouse. |
| Single capture |
Captures a single set of data |
| Continuous capture display: |
Unconditionally Continuous: Trigger forced internally and display updated with regular intervals.
Conditionally Continuous: Display updated when a trigger condition is detected. Any of the above trigger conditions may be set as described above. Updates the display each time a specified edge or pattern trigger condition is met |
| Cursors: |
T-trigger line: Indicates trigger position.
P-trigger line: Indicates trigger condition-1 position in case of condition-1 THEN condition-2 setting.
S-Cursor line: Indicates data search positions.
X, Y and Z-cursor: For time measurements.
The values of the X and Y cursor lines, where they intersect the signal lines are shown. |
| Time measurements: |
The time differences between any two cursor lines or trigger line may be displayed. Take measurement easily by clicking mouse on first edge and on second edge to get the time difference. The time difference may also be indicated as a frequency. |
| Edge snapping: |
Cursors (X, Y & Z) snap to signal edges for accurate time measurements. (Especially useful when the display is zoomed out so that one screen pixel represents more than one data sample).
The signal edges to which the cursors lines have been snapped are indicated |
| Pattern search: |
Any channel conditions may be searched for. 1, 0 and don't care conditions specified.
Also repeated search |
| Saving to disk-formats |
Binary data, Hex, Spreadsheet(CSV), Configuration |
| Printed output: |
The timing diagrams, bitmaps, binary and hex data, may be.
Landscape and portrait. |
| Connection to PC |
USB 2.0 High Speed (USB 1.1 compatible) |
| Protocols |
UART/RS232 Monitor/generate signals
( More protocols to be added ) |
| Extension port |
Useful signal inputs/outputs
E.g.: Output trigger to combine with external oscilloscope.
Pattern generator strobes.
UART Rx/Tx busy, fifo empty/full, etc |
| Power requirements: |
2.5W Typical (Capturing data, without Pattern Generator)
3W Typical (including Pattern Generator)
Power consumption is dependent on number of channels used and input frequencies.
Fused aT 1.6A (or less)
Resetable fuse. Remove power for 3 minutes to reset. |
| LA-Gold-16 unit dimensions/weight |
220 g
162mm x 84mm x 26mm |
| Package contents |
The LA-Gold-16 instrument
16 quality SMD test clips with ground leads
User’s manual hardcopy
Software and documentation on CD |
| Hardware self test |
Comprehensive hardware self test capabilities |
| Instrument integration |
Can combine with logic analyzer operations |
| Operating Temperature |
10-40C. (50-105F) |